The tissue in this test was warmed to 37 degree celsius to simulate body temperature in a water bath. The tissue used in these images was chicken breasts.
For the monopolar testing the radiofrequency generator was set to 80 degrees celsius for 80 seconds with a 30 second ramp for both the Nimbus and the standard 18g cannula.
For the bipolar testing the radiofrequency generator was set to 85°C for 150 seconds for 150 seconds with a 30 second ramp for both the Nimbus and the standard 18g cannula.